Multidetector-row computed tomography imaging characteristics of mechanical prosthetic valves
Publication year
2011Source
Journal of Heart Valve Disease, 20, 2, (2011), pp. 216-22ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Radiology
Journal title
Journal of Heart Valve Disease
Volume
vol. 20
Issue
iss. 2
Page start
p. 216
Page end
p. 22
Subject
NCEBP 14: Cardiovascular diseases; ONCOL 5: Aetiology, screening and detection; Medical Imaging - Radboud University Medical CenterAbstract
BACKGROUND AND AIM OF THE STUDY: Electrocardiogram-gated multidetector-row computed tomography (MDCT) imaging may aid in the evaluation of prosthetic valve dysfunction. A pulsatile in vitro model was developed to study the MDCT imaging characteristics of mechanical heart valves (MHVs). METHODS: Bjork-Shiley (BS), St. Jude Medical (SJM), Medtronic-Hall (MH), CarboMedics (CM) and ON-X valves were inserted into an in vitro pulsatile model and scanned using a 64-detector row scanner. The image quality regarding visualization of the leaflets, prosthetic detail and periprosthetic detail, as well as the presence of image artifacts, was scored on a four-point scale. RESULTS: The image quality for the BS valve was scored poor to moderate for all criteria, and was inferior to that seen with the other valves. Leaflet visualization was excellent for all the other valves. The prosthetic detail was good for MH and SJM valves, and excellent for the CM and ON-X valves. Periprosthetic detail was good for MH, SJM and CM valves, and excellent for ON-X valves. Artifacts were moderate for MH and SJM valves, minor for CM, and minimal for ON-X. All differences were shown to be statistically significant (p < 0.001). CONCLUSION: The pulsatile in vitro model is an effective tool to detect differences in the MDCT imaging characteristics of MHVs. The image quality is determined by the prosthesis components and, to a lesser degree, by prosthesis design. Modern carbon-titanium MHVs yield good to excellent image quality on MDCT.
This item appears in the following Collection(s)
- Academic publications [242767]
- Faculty of Medical Sciences [92292]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.