Potential of Lesion-to-Fat Elasticity Ratio Measured by Shear Wave Elastography to Reduce Benign Biopsies in BI-RADS 4 Breast Lesions.
Publication year
2023Author(s)
Source
Journal of Ultrasound in Medicine, 42, 8, (2023), pp. 1729-1736ISSN
Annotation
01 augustus 2023
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Medical Imaging
Journal title
Journal of Ultrasound in Medicine
Volume
vol. 42
Issue
iss. 8
Page start
p. 1729
Page end
p. 1736
Subject
Radboud University Medical Center; Radboudumc 17: Women's cancers Medical ImagingAbstract
OBJECTIVES: We evaluated whether lesion-to-fat ratio measured by shear wave elastography in patients with Breast Imaging Reporting and Data System (BI-RADS) 3 or 4 lesions has the potential to further refine the assessment of B-mode ultrasound alone in breast cancer diagnostics. METHODS: This was a secondary analysis of an international diagnostic multicenter trial (NCT02638935). Data from 1288 women with breast lesions categorized as BI-RADS 3 and 4a-c by conventional B-mode ultrasound were analyzed, whereby the focus was placed on differentiating lesions categorized as BI-RADS 3 and BI-RADS 4a. All women underwent shear wave elastography and histopathologic evaluation functioning as reference standard. Reduction of benign biopsies as well as the number of missed malignancies after reclassification using lesion-to-fat ratio measured by shear wave elastography were evaluated. RESULTS: Breast cancer was diagnosed in 368 (28.6%) of 1288 lesions. The assessment with conventional B-mode ultrasound resulted in 53.8% (495 of 1288) pathologically benign lesions categorized as BI-RADS 4 and therefore false positives as well as in 1.39% (6 of 431) undetected malignancies categorized as BI-RADS 3. Additional lesion-to-fat ratio in BI-RADS 4a lesions with a cutoff value of 1.85 resulted in 30.11% biopsies of benign lesions which correspond to a reduction of 44.04% of false positives. CONCLUSIONS: Adding lesion-to-fat ratio measured by shear wave elastography to conventional B-mode ultrasound in BI-RADS 4a breast lesions could help reduce the number of benign biopsies by 44.04%. At the same time, however, 1.98% of malignancies were missed, which would still be in line with American College of Radiology BI-RADS 3 definition of <2% of undetected malignancies.
This item appears in the following Collection(s)
- Academic publications [246515]
- Electronic publications [134102]
- Faculty of Medical Sciences [93308]
- Open Access publications [107627]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.