Publication year
2002Source
Clinical & Laboratory Haematology, 24, 6, (2002), pp. 343-5ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Clinical Chemistry
Journal title
Clinical & Laboratory Haematology
Volume
vol. 24
Issue
iss. 6
Page start
p. 343
Page end
p. 5
Subject
Innovation and Quality assurance in laboratory medicine; Ontwikkeling en kwaliteitsborging in de laboratoriumgeneeskundeAbstract
Nucleated red blood cells (NRBC) in blood samples interfere with the white blood cell (WBC) count on many types of automated haematology analysers. This makes it necessary to correct the WBC count by counting NRBC microscopically. This report describes the evaluation of two analysers, the Cell-Dyn 4000 and the Sysmex XE-2100, which use new techniques to recognize and enumerate NRBC. We conclude that both the Cell-Dyn 4000 and the Sysmex XE-2100 give an accurate WBC count in the presence of NRBC. Furthermore, they can enumerate NRBC correctly when compared with microscopic observation.