Variation in arterial input function in a large multicenter computed tomography perfusion study
Publication year
2021Source
European Radiology, 31, 11, (2021), pp. 8317-8325ISSN
Publication type
Article / Letter to editor

Display more detailsDisplay less details
Organization
Medical Imaging
Journal title
European Radiology
Volume
vol. 31
Issue
iss. 11
Page start
p. 8317
Page end
p. 8325
Subject
Radboudumc 9: Rare cancers RIHS: Radboud Institute for Health SciencesAbstract
OBJECTIVES: To report the variation in computed tomography perfusion (CTP) arterial input function (AIF) in a multicenter stroke study and to assess the impact this has on CTP results. METHODS: CTP datasets from 14 different centers were included from the DUtch acute STroke (DUST) study. The AIF was taken as a direct measure to characterize contrast bolus injection. Statistical analysis was applied to evaluate differences in amplitude, area under the curve (AUC), bolus arrival time (BAT), and time to peak (TTP). To assess the clinical relevance of differences in AIF, CTP acquisitions were simulated with a realistic anthropomorphic digital phantom. Perfusion parameters were extracted by CTP analysis using commercial software (IntelliSpace Portal (ISP), version 10.1) as well as an in-house method based on block-circulant singular value decomposition (bSVD). RESULTS: A total of 1422 CTP datasets were included, ranging from 6 to 322 included patients per center. The measured values of the parameters used to characterize the AIF differed significantly with approximate interquartile ranges of 200-750 HU for the amplitude, 2500-10,000 HU.s for the AUC, 0-17 s for the BAT, and 10-26 s for the TTP. Mean infarct volumes of the phantom were significantly different between centers for both methods of perfusion analysis. CONCLUSIONS: Although guidelines for the acquisition protocol are often provided for centers participating in a multicenter study, contrast medium injection protocols still vary. The resulting volumetric differences in infarct core and penumbra may impact clinical decision making in stroke diagnosis. KEY POINTS: * The contrast medium injection protocol may be different between stroke centers participating in a harmonized multicenter study. * The contrast medium injection protocol influences the results of X-ray computed tomography perfusion imaging. * The contrast medium injection protocol can impact stroke diagnosis and patient selection for treatment.
This item appears in the following Collection(s)
- Academic publications [232297]
- Electronic publications [115532]
- Faculty of Medical Sciences [89118]
- Open Access publications [82813]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.