How does image quality affect radiologists' perceived ability for image interpretation and lesion detection in digital mammography?
Publication year
2021Source
European Radiology, 31, 7, (2021), pp. 5335-5343ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Medical Imaging
Health Evidence
Journal title
European Radiology
Volume
vol. 31
Issue
iss. 7
Page start
p. 5335
Page end
p. 5343
Subject
Radboudumc 17: Women's cancers RIHS: Radboud Institute for Health Sciences; Health Evidence - Radboud University Medical Center; Medical Imaging - Radboud University Medical CenterAbstract
OBJECTIVES: To study how radiologists' perceived ability to interpret digital mammography (DM) images is affected by decreases in image quality. METHODS: One view from 45 DM cases (including 30 cancers) was degraded to six levels each of two acquisition-related issues (lower spatial resolution and increased quantum noise) and three post-processing-related issues (lower and higher contrast and increased correlated noise) seen during clinical evaluation of DM systems. The images were shown to fifteen breast screening radiologists from five countries. Aware of lesion location, the radiologists selected the most-degraded mammogram (indexed from 1 (reference) to 7 (most degraded)) they still felt was acceptable for interpretation. The median selected index, per degradation type, was calculated separately for calcification and soft tissue (including normal) cases. Using the two-sided, non-parametric Mann-Whitney test, the median indices for each case and degradation type were compared. RESULTS: Radiologists were not tolerant to increases (medians: 1.5 (calcifications) and 2 (soft tissue)) or decreases (median: 2, for both types) in contrast, but were more tolerant to correlated noise (median: 3, for both types). Increases in quantum noise were tolerated more for calcifications than for soft tissue cases (medians: 3 vs. 4, p = 0.02). Spatial resolution losses were considered less acceptable for calcification detection than for soft tissue cases (medians: 3.5 vs. 5, p = 0.001). CONCLUSIONS: Perceived ability of radiologists for image interpretation in DM was affected not only by image acquisition-related issues but also by image post-processing issues, and some of those issues affected calcification cases more than soft tissue cases. KEY POINTS: * Lower spatial resolution and increased quantum noise affected the radiologists' perceived ability to interpret calcification cases more than soft tissue lesion or normal cases. * Post-acquisition image processing-related effects, not only image acquisition-related effects, also impact the perceived ability of radiologists to interpret images and detect lesions. * In addition to current practices, post-acquisition image processing-related effects need to also be considered during the testing and evaluation of digital mammography systems.
This item appears in the following Collection(s)
- Academic publications [244280]
- Electronic publications [131328]
- Faculty of Medical Sciences [92906]
- Open Access publications [105276]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.