Validation of a Web-Based Planning Tool for Percutaneous Cryoablation of Renal Tumors
Publication year
2020Source
Cardiovascular and Interventional Radiology, 43, 11, (2020), pp. 1661-1670ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Medical Imaging
Radboudumc Extern
Journal title
Cardiovascular and Interventional Radiology
Volume
vol. 43
Issue
iss. 11
Page start
p. 1661
Page end
p. 1670
Subject
Radboudumc 0: Other Research RIHS: Radboud Institute for Health Sciences; Radboudumc 15: Urological cancers RIHS: Radboud Institute for Health Sciences; Medical Imaging - Radboud University Medical CenterAbstract
PURPOSE: To validate a simulation environment for virtual planning of percutaneous cryoablation of renal tumors. MATERIALS AND METHODS: Prospectively collected data from 19 MR-guided procedures were used for validation of the simulation model. Volumetric overlap of the simulated ablation zone volume (Sigma) and the segmented ablation zone volume (S; assessed on 1-month follow-up scan) was quantified. Validation metrics were DICE Similarity Coefficient (DSC; the ratio between twice the overlapping volume of both ablation zones divided by the sum of both ablation zone volumes), target overlap (the ratio between the overlapping volume of both ablation zones to the volume of S; low ratio means S is underestimated), and positive predictive value (the ratio between the overlapping volume of both ablation zones to the volume of Sigma; low ratio means S is overestimated). Values were between 0 (no alignment) and 1 (perfect alignment), a value > 0.7 is considered good. RESULTS: Mean volumes of S and Sigma were 14.8 cm(3) (+/- 9.9) and 26.7 cm(3) (+/- 15.0), respectively. Mean DSC value was 0.63 (+/- 0.2), and >/= 0.7 in 9 cases (47%). Mean target overlap and positive predictive value were 0.88 (+/- 0.11) and 0.53 (+/- 0.24), respectively. In 17 cases (89%), target overlap was >/= 0.7; positive predictive value was >/= 0.7 in 4 cases (21%) and < 0.6 in 13 cases (68%). This indicates S is overestimated in the majority of cases. CONCLUSION: The validation results showed a tendency of the simulation model to overestimate the ablation effect. Model adjustments are necessary to make it suitable for clinical use.
This item appears in the following Collection(s)
- Academic publications [246326]
- Electronic publications [133958]
- Faculty of Medical Sciences [93294]
- Open Access publications [107441]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.