Physical evaluation of an ultra-high-resolution CT scanner
Publication year
2020Source
European Radiology, 30, 5, (2020), pp. 2552-2560ISSN
Publication type
Article / Letter to editor

Display more detailsDisplay less details
Organization
Medical Imaging
Journal title
European Radiology
Volume
vol. 30
Issue
iss. 5
Page start
p. 2552
Page end
p. 2560
Subject
Radboudumc 0: Other Research RIHS: Radboud Institute for Health Sciences; Radboudumc 16: Vascular damage RIHS: Radboud Institute for Health Sciences; Radboudumc 17: Women's cancers RIHS: Radboud Institute for Health SciencesAbstract
OBJECTIVES: To evaluate the technical performance of an ultra-high-resolution CT (UHRCT) system. METHODS: The physico-technical capabilities of a novel commercial UHRCT system were assessed and compared with those of a current-generation multi-detector (MDCT) system. The super-high-resolution (SHR) mode of the system uses 0.25 mm (at isocentre) detector elements (dels) in the in-plane and longitudinal directions, while the high-resolution (HR) mode bins two dels in the longitudinal direction. The normal-resolution (NR) mode bins dels 2 x 2, resulting in a del-size equivalent to that of the MDCT system. In general, standard procedures and phantoms were used to perform these assessments. RESULTS: The UHRCT MTF (10% MTF 4.1 lp/mm) is twice as high as that of the MDCT (10% MTF 1.9 lp/mm), which is comparable to the MTF in the NR mode (10% MTF 1.7 lp/mm). The width of the slice sensitivity profile in the SHR mode (FWHM 0.45 mm) is about 60% of that of the MDCT (FWHM 0.77 mm). Uniformity and CT numbers are within the expected range. Noise in the high-resolution modes has a higher magnitude and higher frequency components compared with MDCT. Low-contrast visibility is lower for the NR, HR and SHR modes compared with MDCT, but about a 14%, for NR, and 23%, for HR and SHR, dose increase gives the same results. CONCLUSIONS: HR and SHR mode scanning results in double the spatial resolution, with about a 23% increase in dose required to achieve the same low-contrast detectability. KEY POINTS: * Resolution on UHRCT is up to twice as high as for the tested MDCT. * With abdominal settings, UHRCT needs higher dose for the same low-contrast detectability as MDCT, but dose is still below achievable levels as defined by current diagnostic reference levels. * The UHRCT system used in normal-resolution mode yields comparable resolution and noise characteristics as the MDCT system.
This item appears in the following Collection(s)
- Academic publications [227409]
- Electronic publications [108578]
- Faculty of Medical Sciences [86883]
- Open Access publications [77814]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.