Improvement of image quality on low-dose dynamic myocardial perfusion computed tomography with a novel 4-dimensional similarity filter
Publication year
2020Source
Medicine (Baltimore), 99, 26, (2020), article e20804ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Medical Imaging
Journal title
Medicine (Baltimore)
Volume
vol. 99
Issue
iss. 26
Subject
Radboudumc 15: Urological cancers RIHS: Radboud Institute for Health Sciences; Radboudumc 16: Vascular damage RIHS: Radboud Institute for Health Sciences; Medical Imaging - Radboud University Medical CenterAbstract
The aim of this study was to evaluate the effect of a novel 4-dimensional similarity filter (4DSF) on quantitative and qualitative parameters of low-dose dynamic myocardial computed tomography perfusion (CTP) images.In this retrospective study, medical records of 32 patients with suspected or known coronary artery disease who underwent dynamic myocardial CTP at 80 kV were included. The 4DSF reduces noise by averaging voxels that have similar dynamic behavior after adaptive iterative dose reduction 3D (AIDR3D) and deformable image registration were applied. Qualitative (artefact, contour sharpness, and myocardial homogeneity [1 = poor; 2 = intermediate; 3 = good]) and quantitative measurement (standard deviation [SD] and signal-to-noise ratio [SNR]) were compared between the 4DSF and AIDR3D. Contrast-to-noise ratio (CNR) between ischemic and normal remote myocardium was also assessed using myocardial perfusion magnetic resonance imaging as the reference standard in seven patients.The 4DSF was successfully applied to all the images. Improvement in subjective image quality yielded by 4DSF was higher than that yielded by AIDR3D (homogeneity, 1.0 [3 vs 2]; artefact, 1.5 [3 vs 1.5]; P < .001) in all patients. The 4DSF significantly decreased the SD by 59% (AIDR3D vs 4DSF: 33.5 ± 0.4 vs 13.8 ± 0.4, P < .001), increased the SNR by 134% (AIDR3D vs 4DSF: 4.4 ± 0.2 vs 10.3 ± 0.2, P < .001), and increased the CNR by 131% (AIDR3D vs 4DSF: 1.6 ± 0.2 vs 3.7 ± 0.2, P < .001).The 4DSF improved the qualitative and quantitative parameters of low-dose dynamic myocardial CTP images.
This item appears in the following Collection(s)
- Academic publications [246626]
- Electronic publications [134196]
- Faculty of Medical Sciences [93367]
- Open Access publications [107722]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.