Experimental and computational analysis of micromotions of an uncemented femoral knee implant using elastic and plastic bone material models
Publication year
2017Source
Journal of Biomechanics, 61, (2017), pp. 137-143ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Orthopaedics
Journal title
Journal of Biomechanics
Volume
vol. 61
Page start
p. 137
Page end
p. 143
Subject
Radboudumc 10: Reconstructive and regenerative medicine RIHS: Radboud Institute for Health Sciences; Orthopaedics - Radboud University Medical CenterAbstract
It is essential to calculate micromotions at the bone-implant interface of an uncemented femoral total knee replacement (TKR) using a reliable computational model. In the current study, experimental measurements of micromotions were compared with predicted micromotions by Finite Element Analysis (FEA) using two bone material models: linear elastic and post-yield material behavior, while an actual range of interference fit was simulated. The primary aim was to investigate whether a plasticity model is essential in order to calculate realistic micromotions. Additionally, experimental bone damage at the interface was compared with the FEA simulated range. TKR surgical cuts were applied to five cadaveric femora and micro- and clinical CT- scans of these un-implanted specimens were made to extract geometrical and material properties, respectively. Micromotions at the interface were measured using digital image correlation. Cadaver-specific FEA models were created based on the experimental set-up. The average experimental micromotion of all specimens was 53.1+/-42.3microm (mean+/-standard deviation (SD)), which was significantly higher than the micromotions predicted by both models, using either the plastic or elastic material model (26.5+/-23.9microm and 10.1+/-10.1microm, respectively; p-value<0.001 for both material models). The difference between the two material models was also significant (p-value<0.001). The predicted damage had a magnitude and distribution which was comparable to the experimental bone damage. We conclude that, although the plastic model could not fully predict the micro motions, it is more suitable for pre-clinical assessment of a press-fit TKR implant than using an elastic bone model.
This item appears in the following Collection(s)
- Academic publications [242527]
- Faculty of Medical Sciences [92283]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.