Interference of daratumumab in monitoring multiple myeloma patients using serum immunofixation electrophoresis can be abrogated using the daratumumab IFE reflex assay (DIRA)
Publication year
2016Source
Clinical Chemistry and Laboratory Medicine, 54, 6, (2016), pp. 1105-9ISSN
Publication type
Article / Letter to editor
Display more detailsDisplay less details
Organization
Haematology
Paediatrics - OUD tm 2017
Laboratory Medicine
Journal title
Clinical Chemistry and Laboratory Medicine
Volume
vol. 54
Issue
iss. 6
Page start
p. 1105
Page end
p. 9
Subject
Radboudumc 5: Inflammatory diseases RIMLS: Radboud Institute for Molecular Life Sciences; Radboudumc 9: Rare cancers RIHS: Radboud Institute for Health SciencesAbstract
Daratumumab is a fully human anti-CD38 IgG1-kappa monoclonal antibody (mAb) currently being evaluated in several Phase 2 and 3 clinical studies for the treatment of multiple myeloma (MM). In this clinical case study we demonstrate that daratumumab can be detected as an individual monoclonal band in serum immunofixation electrophoresis (IFE). M-protein follow-up by IFE is part of the International Myeloma Working Group (IMWG) criteria to assess treatment response. Therefore, it is crucial that the daratumumab band is not confused with the endogenous M-protein of the patient during IFE interpretation. Moreover, a significant number of IgG-kappa M-proteins co-migrate with daratumumab. Co-migration introduces a bias in the M-protein quantification since pharmacokinetic studies show that daratumumab peak plasma concentrations reach up to 1 g/L. More importantly, co-migration can mask clearance of the M-protein by IFE which is necessary for classification of complete response by IMWG criteria (negative serum IFE). For optimal M-protein monitoring the laboratory specialist needs to be informed when patients receive daratumumab, and it is essential that the laboratory specialist is aware that a slow migrating band in the gamma-region in those patients may be derived from the daratumumab. A daratumumab specific IFE reflex assay (DIRA) has been developed and can be utilized to abrogate interference. The here described mAb interference is not limited to daratumumab, and as therapeutic antibodies gain approval and enter into common clinical practice, laboratory specialists will need additional processes to characterize IFE interference and distinguish endogenous M-protein from therapeutic antibodies.
This item appears in the following Collection(s)
- Academic publications [246216]
- Electronic publications [133894]
- Faculty of Medical Sciences [93266]
- Open Access publications [107414]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.