Cardiac output measurement with transpulmonary ultrasound dilution is feasible in the presence of a left-to-right shunt: a validation study in lambs.

Fulltext:
110488.pdf
Embargo:
until further notice
Size:
457.3Kb
Format:
PDF
Description:
Publisher’s version
Publication year
2012Source
British Journal of Anaesthesia, 108, 3, (2012), pp. 409-16ISSN
Annotation
01 maart 2012
Publication type
Article / Letter to editor

Display more detailsDisplay less details
Organization
Paediatrics - OUD tm 2017
Journal title
British Journal of Anaesthesia
Volume
vol. 108
Issue
iss. 3
Page start
p. 409
Page end
p. 16
Subject
IGMD 1: Functional imagingAbstract
BACKGROUND: Cardiac output (CO) monitoring remains complex in newborns as most of the current technologies fail to accurately measure systemic blood flow in the presence of shunts. We validated CO measurements using transpulmonary ultrasound dilution (TPUD) in a neonatal lamb model with a left-to-right shunt. METHODS: Regular arterial and central venous catheters were inserted into seven lambs (3.5-8.3 kg). A surgically constructed left-to-right aorto-pulmonary Gore-Tex((R)) shunt was intermittently opened and closed, while CO was manipulated by creating haemorrhagic hypotension. CO measurements with TPUD (COtpud) were compared with those obtained by an ultrasonic transit-time flow probe positioned around the main pulmonary artery (COufp). RESULTS: We performed 72 sessions of three paired CO measurements. The mean COufp was 1.00 litre min(-1) (range 0.47-1.75 litre min(-1)) and mean COtpud 1.05 litre min(-1) (range 0.54-1.87 litre min(-1)). With an open shunt, the mean Qp/Qs ratio was 1.8 (range 1.3-2.6). A comparison between COufp and COtpud showed a mean bias (sd) of 0.03 (0.09) and 0.07 (0.10) litre min(-1), respectively, for measurements with a closed and an open shunt. The percentage error was 18% and 20% for measurements with a closed and an open shunt. Polar plot analysis showed good trending ability for both closed and open shunt groups. CONCLUSIONS: TPUD is a reliable technology to measure CO in the presence of a left-to-right shunt.
This item appears in the following Collection(s)
- Academic publications [234412]
- Electronic publications [117392]
- Faculty of Medical Sciences [89250]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.