Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT.
Publication year
2012Source
European Radiology, 22, 6, (2012), pp. 1271-7ISSN
Annotation
01 juni 2012
Publication type
Article / Letter to editor

Display more detailsDisplay less details
Organization
Radiology
Journal title
European Radiology
Volume
vol. 22
Issue
iss. 6
Page start
p. 1271
Page end
p. 7
Subject
ONCOL 5: Aetiology, screening and detection; NCEBP 14: Cardiovascular diseases ONCOL 5: Aetiology, screening and detectionAbstract
OBJECTIVES: Multidetector computed tomography (MDCT) has diagnostic value for the evaluation of prosthetic heart valve (PHV) dysfunction but it is hampered by artefacts. We hypothesised that image acquisition using prospective triggering instead of retrospective gating would reduce artefacts related to pulsating PHV. METHODS: In a pulsatile in vitro model, a mono- and bileaflet PHV were imaged using 256 MDCT at 60, 75 and 90 beats per minute (BPM) with either retrospective gating (120 kV, 600 mAs, pitch 0.2, CTDI(vol) 39.8 mGy) or prospective triggering (120 kV, 200 mAs, CTDI(vol) 13.3 mGy). Two thresholds (>175 and <-45HU), derived from the density of surrounding structures, were used for quantification of hyper- and hypodense artefacts. Image noise and artefacts were compared between protocols. RESULTS: Prospective triggering reduced hyperdense artefacts for both valves at every BPM (P = 0.001 all comparisons). Hypodense artefacts were reduced for the monoleaflet valve at 60 (P = 0.009), 75 (P = 0.016) and 90 BPM (P = 0.001), and for the bileaflet valves at 60 (P = 0.001), 90 (P = 0.001) but not at 75 BPM (P = 0.6). Prospective triggering reduced image noise at 60 (P = 0.001) and 75 (P < 0.03) but not at 90 BPM. CONCLUSIONS: Compared with retrospective gating, prospective triggering reduced most artefacts related to pulsating PHV in vitro. KEY POINTS: * Computed tomographic images are often degraded by prosthetic heart valve-induced artefacts * Prospective triggering reduces prosthetic heart valve-induced artefacts in vitro * Artefact reduction at 90 beats per minute occurs without image noise reduction * Prospective triggering may improve CT image quality of moving hyperdense structures.
This item appears in the following Collection(s)
- Academic publications [229289]
- Faculty of Medical Sciences [87821]
Upload full text
Use your RU credentials (u/z-number and password) to log in with SURFconext to upload a file for processing by the repository team.